Measurement optics for XL-80 Laser

The same type of optics can be used for the XL-80 system, as the ML10. All measurements are interferometric and as a result of this, they use the traceable international standard wavelength of laser light.

The XL-80 optic system has various benefits that every user can enjoy. For example, it is interchangeable – meaning that it enables easy exchange for different measurements, without having to re-align the laser. Secondly, this measurement system also delivers convenience to the user because the moving optics have no cables to drag or snag. Lastly, this measurement system has quick acclimatization; some competitor systems

Please note: For rotary axis calibration, please see the RX10 rotary axis calibrator section. Also, for other optical accessories to assist in set-up and measurement, please see the optics accessories section. use stainless steel optic housings and this system does not. Instead, it uses Renishaw's aluminum optics, which has a higher thermal conductivity level.

LINEAR MEASUREMENT OPTICS KIT

Linear measurement is the most common form of measurement performed on a CMM machine.

Each linear measurement optics kit contains the following:

  • 1 linear interferometer
  • 2 linear retroreflectors and clamp screws
  • 2 alignment targets

Part number: A-8003-0440 Linear measurement optics kit.

Specification

Standard range 0 m - 40 m (0 in - 1600 in)
System accuracy (when used with XL-80 system) ±0.5 ppm (parts per million)
(when used with ML10 Gold Standard) ±0.7 ppm
(when used with ML10, pre-Gold Standard) ±1.1 ppm
Resolution 0.001 µm (0.1 µin)
Maximum velocity 4 m/s (9600 in/min)
Velocity measurement accuracy ±0.05%

For linear measurements over 40 m you will require the long range linear accessory kit

ANGULAR MEASUREMENT OPTICS KIT (STANDARD)

These optics can measure maximum angular deflections of up to ±10° with resolution of 0.01 arc sec. As a result, angular pitch and yaw measurements can be taken.

Each angular measurement kit contains the following:

  • 1 angular retroreflector
  • 1 angular interferometer
  • 2 alignment targets
Axial range 0 m - 15 m (0 in - 590 in)
Angular measurement range ±175 mm/m (±10°)
Angular accuracy ±0.2% ±0.5 ±0.1M µm/m
(±0.2% ±0.1 ±0.007F arc sec)
Resolution 0.1 µm/m (0.01 arc sec)

Where:
% = percentage of displayed value
M = measurement distance in metres
F = measurement distance in feet

ANGULAR MEASUREMENT OPTICS KIT (CALIBRATED)

These optics allow angular pitch and yaw measurements to be made. They also allow a lower measurement uncertainty than with standard optics.

Each calibrated angular measurement kit contains the following:

  • 1 angular retroreflector
  • 1 angular interferometer
  • 2 alignment targets
  • calibration certificate

Part number: A-8003-2532 Angular measurement optics – calibrated

Each retroreflector is supplied with a calibration certificate and marked with a calibration factor.

Specification as for standard angular optics but angular accuracy term is reduced from ±0.2% to ±0.04%, when the calibration factor is entered into LaserXL software.

These particular optics are available with steel housing, for greater thermal stability.

LINEAR AND ANGULAR COMBINATION KIT

This optics combination kit is a cost saving alternative for users who wish to perform both linear and angular measurements.

Each combination kit contains:

  • 1 linear retroreflector and clamp screws
  • 1 angular retroreflector
  • 1 angular interferometer
  • 3 alignment targets

Part number: A-8003-2774 Linear and angular optics combination kit

STRAIGHTNESS MEASUREMENT OPTICS KIT: 0.1 - 4 M AXIS LENGTH

These optics allowaxis parallelism to be measured. They also allow the measurement of horizontal straightness error in linear axes up to 4 m in length to be made.

Each short range straightness kit contains the following:

  • 1 straightness interferometer
  • 1 straightness reflector

Part number: A-8003-0443 Straightness optics - short range

Specification

Axial range 0.1 m - 4 m (4 in - 160 in)
Straightness measurement range ±2.5 mm (±0.1 in)
Accuracy ±0.5% ±0.5 ±0.15M2 µm

(±0.5% ±20 ±0.5F2 µin)

Resolution 0.01 µm (1 µin)

Where:
% = percentage of displayed value
M = measuring distance in metres
F = measuring distance in feet

May require Universal (straightness) shutter or straightness accessory kit, depending upon application

STRAIGHTNESS MEASUREMENT OPTICS KIT: 1 - 30 M AXIS LENGTH

These optics can be used to measure the horizontal straightness error in the linear axes, from 1 m up to 30 m in length. It must be used with the XL-80 laser system. This kit also enables measurement for the parallelism of axes.

Each long range straightness kit contains the following:

  • 1 straightness interferometer
  • 1 straightness retroreflector
  • 2 alignment targets

Part number: A-8003-0444 Straightness optics - long range

Specification

Axial range 1 m - 30 m (40 in - 1200 in)
Straightness measurement range ±2.5 mm (± 0.1 in)
Accuracy* ±2.5% ±5 ±0.015M2 µm

(±2.5% ±200 ±0.05F2 µin)

Resolution 0.1 µm (10 µin)

Where:
% = percentage of displayed value
M = measuring distance in metres
F = measuring distance in feet

* subject to environmental conditions

May require Universal (straightness) shutter or straightness accessory kit, depending upon application.

SQUARENESS MEASUREMENT OPTICS

By comparing their straightness slope values which are referenced by the optical square, these optics determine the out-of-squareness of two nominally orthogonal axes.

Please note: Astraightness measurement optics kit and straightness accessory kit are also required to perform squareness measurements.

Each squareness measurement optics kit contains the following:

  • 1 optical square and clamp screws
  • 1 bracket (for adjustable turning mirror)

Part number: A-8003-0665 Squareness measurement optics

Specification

Range ±3/M mm/m (±2000/F arc sec)
Resolution 0.01 µm/m (0.01 arc sec)

Where:
M = measurement distance in metres
F = measurement distance in feet

FLATNESS MEASUREMENT KIT

The flatness measurement kit not only rotates horizontally, but can also tilt to allow horizontal and vertical adjustment of the laser beam.

Please note: Angular optics are also required to perform flatness measurements.

Each flatness measurement kit contains the following:

  • 2 flatness mirrors
  • 1 50 mm flatness base
  • 1 100 mm flatness base
  • 1 150 mm flatness base

Part number: A-8003-0442 Flatness measurement kit

Specification

Axial range 0 m - 15 m
Flatness measurement range ±1.5 mm (±0.06 in)
Accuracy ±0.6%* ±0.02M2 µm

(±0.6%* ±0.08F2 µin)

Resolution 0.01 µm (1 µin)

Where:
% = percentage of calculated flatness
M = length of diagonal in metres
F = length of diagonal in feet

Get in Touch

Please fill out the form below and we will be in touch as soon as possible, thank you!
linkedin facebook pinterest youtube rss twitter instagram facebook-blank rss-blank linkedin-blank pinterest youtube twitter instagram